OC-SENSOR IO Fully automated immunochemical fecal test for detecting gastrointestinal bleeding associated with disorders such as colorectal cancer, polyps and diverticulitis.
Reagents
Product Name | Description | Pack Size |
---|---|---|
V-PZ25 | OC-Auto Sampling bottle 3 (with barcode) | 100 pcs |
V-PH18 | OC-Auto 3 Latex Regeant | 2 x 7 mL |
V-PH46 | OC-Auto 3 Buffer | 200 mL |
V-PH52 | OC- Calibrator 2 Kit | 1 x 3 mL |
V-PH53 | OC-Control LV1 (low) | 2 x 5 mL |
V-PH54 | OC-Sensor Control LV2 (High) | 2 x 5 mL |
Principle | Latex agglunitination immuno-turbidimetry |
---|---|
Method | One step rate essay |
Working Speed | 88 Tests/hour |
Sample Rack | 20 Samples : 5 samples x 4 racks |
Reaction Cuvette | Disposable acrylic cuvette 10 serial cuvette, Product name ‘DISPO-10’ |
Sampling System | Drawn up by sampling nozzle |
Reageant Dispense | Drawn up by regeant nozzle |
Mixing System | By spinning mixer |
Thermostatic System | 37°C Block Heater |
Light Source | L.E.D ( wavelength 660nm ) |
Light Detector | Silicon Photodiode |
Data Input | Color LCD Touch Panel ( 5.7 inches ) |
Memory Capacity | Sample: 5000 tests, QC: 1000 tests, Error: 1000 %tests |
Data Output | Built-in thermal printer, RS-232C, USB |
Barcodes | Sample barcode, rack barcode |
Power Required | AC100 – 240V, 50/60Hz, 150 VA |
Mixing System | By spinning mixer |
Dimension | W360 x D560 x H425 (mm) / 35kg |